7
7-20 z.one ultra and z.one ultra sp OPERATOR MANUAL
Figure 7-14. Menu Access - Compounding Function
To Activate Image Compounding (SmartCart/SmartCart sp only)
1. During B-Mode, press the Compounding ON/OFF softkey.
2. Alternatively, press the Frequency button to quickly cycle through the
Frequency and Compounding choices.
a. Look at the Data Display to see the Frequency or Compounding
schemes.
To Activate Image Compounding (SmartCart/SmartCart sp & Scan
Engine)
1. Press the Tab button (Scan Engine) or Menu button (SmartCart/SmartCart
sp
), and the Imaging menu will appear.
2. Highlight the Compounding menu item.
3. Press the Select button to activate this function.
To Deactivate Image Compounding (SmartCart/SmartCart sp & Scan
Engine)
1. After selecting (highlighting) the Compounding menu item, press the Select
button again to turn off this function.
Needle Viz
Preset (Linear
array
transducers)
1. Press Exam Type button: Select Small Parts or Vascular
2. Select Needle Viz on next menu.
3. The B Steer softkey displays, showing B Steer –10 (default for preset).
4. Turn B Steer softkey to cycle through options. B Steer (0°); B Steer 10
(10° right); or B Steer –10 (10° left).
NOTE: L8-3 transducer B Steer options: B Steer (0°), B Steer 15 (15° right), B Steer
–15 (15° left). See
page 7-26 for more on B Steer.
B-Mode
Controls
NOTE: To access B-Mode, see page 7-14.