9-48 z.one ultra and z.one ultra sp OPERATOR MANUAL
Enter, Enter/Select, or Select. Additionally, Autotrace Calc results can only be
entered from a frozen strip display (not 'live' or Cine Play).
While Auto-Dop Trace is active, the Doppler envelope will be outlined in
green. When you freeze the strip, Automeasure numbers are displayed at the
bottom right side of the image.
When the image is frozen, vertical BLUE line graphics delimiters will appear
on the PW strip. A small yellow arrow will appear at the top to mark the peak
systolic velocity. When you have multiple cardiac cycles displayed, a small
yellow arrow will also be displayed at the bottom to mark the peak end
diastolic.
The placement of the delimiters is determined by how many heart cycles were
chosen in the N Cycles selection of the Auto Params menu. The heart cycle
selected for active evaluation can be changed by scrolling the Trackball to the
left or right.
You can manually adjust the position of the delimiters by pressing the Set
button. The first press will allow for adjustment of the Left delimiter; the
second press will allowing adjustment of the Right delimiter. The calculated
Automeasure values will be adjusted accordingly.
Auto-Dop Trace is different from conventional autotrace functions in that the
Auto-Dop Trace process is applied to the spectrum prior to any
post-processing functions (such as gain, dynamic range, edge) being applied.
Manual adjustments to any of these post-processing functions will not affect
the Auto-Dop Trace.
NOTE: Auto-Dop Trace is intended to serve as an adjunct to the diagnostic
process in evaluating blood flow during PW Doppler examinations. When using
the Auto-Dop Trace feature, please evaluate the results to verify that you are in
agreement before committing the values to the Calc Report Package.