13
z.one ultra and z.one ultra sp OPERATOR MANUAL 13-13
Reconciliation For exams not yet sent to PACS. Makes the
patient/study data in all images in a series the
same during network storage. The data is taken
from the most recent image stored.
Required settings or state:
Reconciliation checkbox is checked.
Image Transfer Mode on Store/Print
setup is Exam Completed.
Storage device is mapped to one or more
of Store1/Store2/Print.
Images are stored to media and network
destination via configured Store1/Store2/
Print button.
Network connection is available.
Reconciliation (continued) Other information:
The default is unchecked.
If no network connection, the following
message appears on the screen when
End Exam is pressed:
Storage job not
enqueued because reconciliation is enabled
and no network present
.
No effect when storing from archive.
Object Type
Select the type of ‘object’ that will be sent
over the DICOM network connection to a
PACS storage device.
NOTE: Installation sites with DICOM PACS storage
devices that do not accept clips must be configured
for still image store only.
For SR Private Data, the default is
unchecked. For most users, this is the
correct setting. If box is checked, raw
calculations data in XML format is
appended to structured report files,
increasing file size and requiring more
system memory. This option is not needed
for typical system usage. See
“User
Customizable Worksheets” on page 9-7.
For information about IQ Scan Data and
storing
IQ Scan Data, see page 11-10.
For information about SR Document, see
“DICOM Structured Reports” below.
For information about Key Images, see
“Key Images” on page 11-26.
Release Association Specifies when to release the communication link
between the ultrasound system and the DICOM
storage device.
Table 13-4. DICOM Network Storage Destination
Parameters (Continued)
Parameter Description