z.one ultra and z.one ultra sp OPERATOR MANUAL 9-33
OB Trending data will only be displayed on the graphs when the data are selected
to be plotted for the Clinical (LMP); data are not presented on the graph if
Ultrasound is selected.
Procedures for
OB Calculation
Packages
To Acquire GS Calculation
1. Optimize the B-Mode image of the GS (gestational sac).
2. Press Freeze to open the Freeze menu and freeze the B-Mode image.
3. To view the Calculations menu:
4. The Obstetrics Calculations menu displays on the screen. Highlight GS and
then press Enter. An active caliper displays in the Image area.
5. Use the Trackball to position the active caliper to the desired location within
the image display.
6. Press the Set button to fix the first Distance caliper and activate the second
caliper.
7. To finalize the measurement for the first pair and bring up the first distance
caliper of the second pair:
8. Use the Trackball to position the first Distance caliper of the second pair to the
desired location within the image display.
9. Press the Set button to fix the first Distance caliper of the second pair. This will
activate the second Distance caliper of the second pair.
10. Use the Trackball to position this caliper. Then, to finalize the measurement
for the second pair and bring up the first Distance caliper of the third pair:
11. Press Freeze to unfreeze the current image and obtain your next image. (Note
that the caliper will momentarily disappear.)
SmartCart/SmartCart sp Press Calc button
Scan Engine Select Calc
SmartCart Press Enter
SmartCart sp
Press Enter/Select button. Press Set button to
activate next caliper.
Scan Engine Press Select
SmartCart Press Enter
SmartCart sp
Press Enter/Select button. Press Set button to
activate next caliper.
Scan Engine Press Select