7
z.one ultra and z.one ultra sp OPERATOR MANUAL 7-75
NOTE: Once a Mode key is defined, it will remain so unless the system
configuration is reset to factory default settings.
2. Connect and activate a transducer enabled for Contrast Imaging: P4-1c
(Cardiac) or C4-1 (Abdominal).
3. Select an appropriate Preset for the application.
4. While imaging the desired anatomy, press Mode/Function key you
programmed for Contrast.
•For P4-1c, system enters Contrast mode & displays single image.
•For C4-1, system enters Simul Dual mode with Contrast detection image on
left and non-contrast reference image on right.
5. Once in Contrast mode, system displays Contrast softkey controls:
Figure 7-40. Contrast Imaging Softkeys
NOTE: Rotate Gain softkey to control Contrast image brightness. Rotate 2D/B
button to control reference image brightness.
Contrast Imaging User Controls
Contrast imaging controls are accessed via the Control Panel (all systems),
softkey displays (
SmartCart/SmartCart sp), and Imaging menu options (all
systems). They are described in
Table 7-8.
Compound
ON/OFF
Gain
Simul
[Frame
Rate Hz]
[Power
%]
Stopwatch
START
[STOP]
Stopwatch
RESET
[PHASE]
Table 7-8. Contrast Imaging User Controls
Control SmartCart & SmartCart sp
Softkeys
Simul Softkey (C4-1
transducer only)
Controls whether Dual Simultaneous or Dual (non-simultaneous) mode is
operational. By default, Dual Simultaneous mode is enabled.
Press Simul to transition to Dual mode, displaying only the Contrast detection
image.
Press Enter button to toggle between right and left imaging panes.
Press Simul again to return to Dual Simultaneous mode.