15-8 z.one ultra and z.one ultra sp OPERATOR MANUAL
If either component fails to start-up after these steps, contact ZONARE
Technical Support.
Scan Engine (Undocked) Fails To Power ON
Remove the battery from the Scan Engine, and then press the battery charge
status button (on the label side of the battery) to ensure that a charge of
greater than 25% is available. If the battery charge is insufficient, replace the
battery or dock the
Scan Engine in the Cart and allow time to recharge the
battery.
Disconnect the transducer from the Scan Engine (if attached) and then repeat
the attempt at powering
ON the unit.
Dock the Scan Engine on the Cart and attempt to power ON the System. If the
Cart powers
ON, the problem is in the Scan Engine battery. Replace the
battery.
If the Scan Engine still fails to power ON, contact ZONARE Technical Support.
System
Start-Up
Problems
Scan Engine (Undocked) Fails To Properly Boot
Press the Power button on the Scan Engine to begin System boot.
If the Scan Engine fails to boots properly, remove the transducer and try
again.
If the Scan Engine still fails to boot properly, insert a System software installer
USB Memory Stick into the USB port on the rear of the
Scan Engine, and
perform a system software reinstallation process.
If the Scan Engine still fails to complete the boot operation, contact ZONARE
Technical Support.
Peripheral
Problems
Peripheral(s) Fails To Power ON
Check to ensure that AC power is applied to the Cart, and that the System is
powered
ON.
Check the AC power cord connections to the peripheral.
Press the power button on the peripheral to ensure it is powered on.
If the peripheral fails to power ON after these steps, contact ZONARE
Technical Support.
Color or B/W Printer Fails to Print
Verify that the Freeze button has been pressed to obtain a static image before
pressing the assigned print button (Print, etc).
Check that the printer has paper available.
Check that the printer has an operational color ribbon cartridge installed
(Sony UP-D25MD color printer)
Ensure that the USB cable is properly connected between the peripheral and
the Cart.
If the printer fails to print after these steps, contact ZONARE Technical
Support.