13
z.one ultra and z.one ultra sp OPERATOR MANUAL 13-7
Printers
Only DICOM-compliant printers can be configured using this procedure. Consult a
network administrator to determine if the printer of interest is DICOM-compliant.
NOTE: To set up a local USB printer, see “Setting Up or Adding a Local (USB)
Printer” on page 12-5.
To Configure a DICOM Printer
1. Make sure the System has a live Ethernet connection to the target DICOM
devices (
Figure 13-7).
NOTE: DICOM parameters can be configured while disconnected from the
Ethernet connection, but testing will not be possible.
2. Go to: Setup | System Setup | Archive | DICOM | Printer
3. The Printer Administration screen displays (Figure 13-9).
Station Name
e.g., ZONARE
User-defined name identifying the ultrasound
system. Should be provided by the network
administrator.
TCP Timeout (sec) Maximum duration (seconds) of waiting time
for completion of TCP transfers over the
network. Set according to anticipated network
type/speed.
Queue Timeout (sec) Maximum amount of time a DICOM job will
remain active in the queue before it times out.
Queue Start Type Automatic: Attempts to start processing jobs
are made immediately upon system startup.
Manual: User must manually start the queue.
Attempts to start processing jobs are not made
immediately upon system startup.
Keep Alive Timeout Duration (seconds) that a DICOM connection
will be maintained after completion of the last
transfer.
Association Acceptor Check the Port checkbox to enable other
DICOM devices to initiate verification of
DICOM communication (association) with the
ultrasound system. User must be sure to
specify the correct port number (default is
104).
Default is unchecked.
If both the Port and Storage boxes are
checked, the ultrasound system acts as a
DICOM storage device.
Log Level Defines the level of detail displayed in the log.
Consult your network administrator.
Table 13-2. DICOM General Configuration Parameters (Continued)
Parameter Description