14
14-4 z.one ultra and z.one ultra sp OPERATOR MANUAL
• Capturing of system status to LOG files
• Capturing current image screen and storing as a *.bmp file
• Transferring the contents of the internal LOG directory (via an internet
connection) to
ZONARE’s FTP site
• Checking (over the internet) for availability of Software and Firmware (cart)
Updates from remote
ZONARE FTP site
• Remote (over the Internet) downloadable Software and Firmware (cart)
Updates from the
ZONARE FTP site
Figure 14-2. User Diagnostics Panel
To Access the User Diagnostics Panel
NOTE: The majority of the functions offered in the User Diagnostics Panel involve
the use of an Internet connection between the
SmartCart/SmartCart sp and
ZONARE’s network server for the purpose of uploading or downloading of files.
Performing any of these procedures will require contacting
ZONARE Technical
Support to receive specific information on IP address and log-in used to make the
connection.
PC Board Revision
Information
Software Version
System Serial Number
Transfer Log Directory
Contents in Internal
Archive to ZONARE
FTP Folder
Capture System Log
Files and Store to
Internal Archive
Store a *.BMP screen
image to Internal
Archive
Capture System Log
Files and Store to USB
Memory Stick or other
External Media
SmartCart/SmartCart sp While viewing an imaging display, press the
Service key on the
QWERTY keyboard:
Single short press activates the User
Diagnostics Panel
Long press (1 sec) will Store Current Logs;
system beeps as logs are stored
Press Service key again to return to imaging
Scan Engine Go to: Tools | System Setup | Diagnostics