z.one ultra and z.one ultra sp OPERATOR MANUAL 6-15
Figure 6-17. TEE Transducer - Connector and Pins
WARNING: Immediately replace a transducer that exhibits any damage.
Cleaning &
Disinfecting
TEE
Transducer
NOTE: Do not use any cleaner or disinfectant on the transducer pins. A dry lint-free
cloth can be used to wipe the pins or a lint-free cloth slightly dampened with an
approved disinfectant. Never press the pins with your finger as the oils in your
fingers could interrupt contact.
ZONARE recommends cleaning the transducer using the Wipe Method after each
patient exam. The use of a soft cloth thoroughly moistened with an approved
disinfectant is recommended for cleaning the strain relief, control housing, cable,
and connector.
ZONARE recommends disinfecting the transducer using the Immersion Method
after each patient exam. The use of a soft cloth thoroughly moistened with an
approved disinfectant is recommended for cleaning the strain relief, transducer
housing, cable and connector.
ZONARE also recommends the use of either a
sterile or nonsterile transducer cover/sheath when endocavity transducers are
used.
NOTE: The transducers cannot be sterilized. When sterility is required, use a
sterile transducer cover/sheath and sterile ultrasound/coupling transducer gel.
Follow all sheath manufacturer prescribed techniques.
WARNING: Before cleaning any device, turn off the system and disconnect power
cord from AC power source to avoid electrical shock.
WARNING: Always examine transducers for damage, such as cracks, splitting,
holes, or fluid leaks. If damage is evident, discontinue use of the transducer and
contact
ZONARE Technical Support.
WARNING: Always use protective eyewear and clothing when cleaning or
disinfecting the transducers.
WARNING: Do not allow the disinfectant to come in contact with metal surfaces
(transducer connector). Use a soft cloth and warm soapy water to remove any
disinfectant that remains on metal surfaces.
CAUTION: When a brush is used for cleaning the transducer, use only a soft
brush; coarse/stiff-bristle brushes may cause transducer damage.
Connector pins