7
7-48 z.one ultra and z.one ultra sp OPERATOR MANUAL
Sweep
Speed
SmartCart/
SmartCart sp
NOTE: See “Sweep Speed” on page 10-7 for data on sweep speeds (mm/s) when ECG is active.
Press Menu button. Highlight Sweep Speed with Trackball. Move Trackball left/right (or
press left/right Set buttons) to choose sweep rate selection (Low, Medium, High). Selected
value displays to left of PW-strip.
Disp Format
SmartCart/
SmartCart sp
NOTE: The Display Format size can be stored to a Preset.
Press Menu button. Highlight Disp Format (display format) with Trackball. Move Trackball
left/right (or press left/right Set buttons) to choose preferred display format.
Available display formats:
PW-strip (frozen image) or scroll (live image)
Split-screen (B-Mode image above and PW-strip below), which has options for relative sizes
of B-Mode image and PW-mode strip size
Maps
SmartCart/
SmartCart sp
Press Menu button. Highlight Maps with Trackball. Move Trackball left/right (or press left/
right Set buttons) to choose preferred grey-scale maps (M1, M2, M3, M4) to be used in
spectral Doppler.
Table 7-4. PW Doppler Controls (Continued)
PW Control SmartCart/SmartCart sp Scan Engine
Indicates that this control is also available in retrospective processing.