z.one ultra and z.one ultra sp OPERATOR MANUAL 6-13
To lower temperature, freeze image for a short time or lower Acoustic
Output power
3. MAXIMUM TEMP LIMIT EXCEEDED
Figure 6-15. Temperature Indicator When Max Temp Exceeded
Figure 6-16. Temperature Warning When Max Temp Exceeded
Image will freeze and display Warning in Figure 6-16.
Click OK and disconnect TEE transducer. Wait a few minutes, then
reconnect TEE transducer to MTP.
Press TEE transducer softkey to reactivate transducer.
If TEE transducer is cool enough, system will resume imaging.
If TEE transducer is still too hot, system will display:
Click OK and disconnect TEE transducer. Wait a few minutes, then
reconnect TEE transducer to MTP.
Press TEE transducer softkey to reactivate transducer.
If TEE transducer is cool enough, system will resume imaging.
Examination The actual examination with the P8-3TEE transducer is beyond the scope of this
manual. There are many medical articles and books that thoroughly address this
topic. There are however specific cautions that should be observed.
CAUTIONS
Long-term exposure to ultrasound should be minimized. Although there have
been no confirmed adverse effects produced by diagnostic levels of
ultrasound, unnecessary patient exposure to ultrasound energy should be
avoided, especially in the Doppler mode.