z.one ultra and z.one ultra sp OPERATOR MANUAL 9-35
5. Heart Rate (HR) is active by default. A green vertical bar displays on the
scroll.
6. Use the Trackball to position the green vertical bar at the beginning of a
cardiac deflection. Press Set to lock the position of the bar. A second bar
becomes active.
7. Use the Trackball to position the second vertical bar at the beginning of the
next cardiac deflection (the number of cycles to use is configured on the Calcs
General Configuration screen; see
Figure 9-5 on page 9-5).
8. To lock the second green bar and complete the measurement:
9. The HR measurement result is copied to the Worksheet. The data display in
the Measurements window inverts to black on white.
10. To view the Worksheet:
11. To edit the Worksheet, highlight the value to be edited, then press Del Txt or
use Bksp to overtype replacement text.
12. To review the Report:
Additional Calculation Packages
Examples of the other supported Calculation Packages are described in the
tables that follow. Steps for performing these calculations are similar to those
previously described in the Generic and OB Calculation Package examples. Due
to general similarity, detailed steps will not be repeated for these Calculation
Packages unless there is a significant difference in the measurement protocol
from that already established in the Generic and OB Calculation Packages.
Scan Engine Select Calculations
SmartCart Press Enter
SmartCart sp
Press Enter/Select button
Scan Engine Press Select
SmartCart/SmartCart sp Press Report
Scan Engine Select Report from the Calculations menu
SmartCart/SmartCart sp Press Report
Scan Engine Select Report from the Calculations menu